Vista de pagina 60
Table 5 GeneAmp
®
PCR System 9700 PCR Method
PCR Protocol
Run
Speed
Reaction
Volume
Stage 1 Stage 2 Stage 3
96.0°C 60.0°C 98.0°C 60.0°C 10.0°C
[1]
Standard 20 μL10:00 2:00 0:30 2:00 99:59
1x (Hold) 39x (Cycles) 1x (Hold)
[1]
O
ptional step.
Thermal cycle the Digital PCR 20K Chips
The following materials are required to thermal cycle prepared QuantStudio
®
3D
Digital PCR 20K Chips for use with the QuantStudio
®
3D Digital PCR System.
Prepared QuantStudio
®
3D Digital PCR 20K Chips (from the previous section)
ProFlex
2x Flat PCR System (or Dual Flat Block GeneAmp
®
PCR System 9700)
QuantStudio
®
3D Digital PCR Chip Adapters
QuantStudio
®
3D Digital PCR Thermal Pads
QuantStudio
®
3D Digital PCR Tilt Base
Isopropanol
Gloves, powder-free, nitrile
Cleanroom-grade, low-lint polyester wipes
Materials
required
Chapter 3 Perform the PCR
Thermal cycle the Digital PCR 20K Chips
3
QuantStudio
®
3D Digital PCR System User Guide