Maintenance
Cleaning the chip tray and sample block ................................ 86
Configure the instrument settings ...................................... 88
Returning your QuantStudio
®
3D Instrument for service .................. 89
Replace the instrument fuses ........................................... 90
Update the instrument firmware ....................................... 91
Calibrate the touchscreen .............................................. 93
View the instrument log ............................................... 93
IMPORTANT! This chapter contains all user service procedures for the QuantStudio
®
3D Digital PCR Instrument. Procedures other than those described in this document
must be performed by a qualified Service Engineer.
Cleaning the chip tray and sample block
If a QuantStudio
®
3D Digital PCR 20K Chip leaks during thermal cycling or during a
run, perform the following procedure to eliminate any residual contaminants from
the QuantStudio
®
3D Instrument chip tray or the thermal cycler.
Cotton or nylon swabs and lint-free cloths
Deionized water
Isopropanol
Pipette (100-μL) with pipette tips
Powder-free gloves
Safety glasses
A
Materials
required
86
QuantStudio
®
3D Digital PCR System User Guide